Brief Introduction: Instrumented Impact Testing Machine, also called Oscillogram Impact Testing Machine, is an ideal equipment to conduct dynamic impact analysis of metallic materials. The fracture behavior of metallic materials is normally evaluated under static and quasi-static loading conditions. However, in real life, the components may experience dynamic loading during events such as accidental overload. It is therefore imperative to study and understand the crack resistance behavior of materials during such dynamic loading conditions.
Standards: ISO14556:2000 Steel-Charpy V-notch pendulum impact test-Instrumented test method ASTM E23 – 07ae1 Standard Test Methods for Notched Bar Impact Testing of Metallic Materials GB/T19748-2005;GB/T3808-2002;GB/T229-2007;GB/T18658
Features
- High stiffness load frame & base to guarantee the stability.
- Refined swing rod with high stiffness prevented the shaking of pendulum after test.
- High precise photoelectric encoder to measure the angle. The whole system adopts lot of mature Optoelectronic Technology and Optoelectronic seclusion technology to improve the measuring precise and stability.
- The machine can not only be operated by software, but also by control box. Dual control way improved the flexibility of machine operation to improve the working efficiceny.
- The software can not only examine the energy, return zero, loss of energy, swing circle, but also can display, store, print the testing result through computer
Specifications
JB-W750 | |
Max. impact energy | 500J/750J |
Pendulum moment | 267.9492NM/ 401.9238NM |
Max. impact speed | 5.4~5.58 m/S |
Rising Angle | 150º |
Distence between centers of Pendulum and speciemn | 800mm |
Span specimen plate | 40mm |
Radius of specimen plate round circle | R1-1.5mm |
Radius of striking edge round circle | R2-2.5mm |
Corner dimension of striking edge | 30 degree |
Thickness of impact blade | 16mm |
Min. resolution of computer display | 0.1J |
Outer dimension | 2300x600x1400 mm |
Weight | 1500KG |
Power supply | AC 380V, 50Hz, 3 phases 4 lines, 550W(main motor) |
Size of specimen | 10 x 10 x 55 mm |
10 x 7.5 x 55 mm | |
10 x 5 x 55 mm |